Omnigrad T TST410
Modular RTD thermometer

Omnigrad T TST410
Modular RTD thermometer ©Endress+Hauser

Comprehensive and common temperature measurement technology for almost all branches of industry

The thermometer is mainly used in the chemical industry but also finds its use in other branches. Typical applications are pressure-less systems such as air ducts, flues and pipelines. For protection against high industrial demands the use of a special protective sleeve is recommended. A variety of dimensions offers flexible application possibilities.

  • 利点

    • High degree of flexibility thanks to modular design with standard terminal heads as per DIN EN 50446 and customer-specific immersion lengths

    • High degree of insert compatibility and design as per DIN 43772

    • High accuracy and exceptional long-term stability

  • アプリケーション

    Typical applications for the TST410 sensor are pressure-less systems such as air ducts, flues and pipelines.
    Special care should be given to the maximum allowed temperature surrounding the terminal head, which must be lower than 80 °C.

機能と仕様

戻る

Details

  • 測定原理

    測温抵抗体

  • 特性 / アプリケーション

    メートル法

    モジュラータイプ

    プロセス接続はコンプレッションフィッティング

    ネックなし

    ミニヘッド

  • ターミナルヘッド

    TA20L - アルミニウム - ミニヘッド

  • サーモウェル / 保護管

    無し (サーモウェルとの併用無し)

  • インサート / プローブ

    無機絶縁 (MI)、形状変更可

  • 保護管/インサート外径

    3.0 mm (0.12'')

  • 最大浸漬長 要問合わせ

    最大 10,000.00 mm (393.70'')

  • 保護管/サーモウェルの材質

    1.4401 (SUS316相当)

  • プロセス接続

    コンプレッション フィッティング:

    G1/4"

  • 先端チップ部の形状

    ストレート

  • 測定温度レンジ

    PT 100:

    -50 °C ...400 °C

    (-58 °F ...752 °F)

  • 最大プロセス圧力 (静的)

    20 °Cの時: 4 MPa (580 psi)

  • 精度

    クラス A(IEC 60751基準)

    クラス B(IEC 60751基準)

  • 応答時間

    t50 = 2 秒

    t90 = 4 秒

  • 組込型伝送器

ドキュメント/マニュアル/ソフトウェア

アクセサリ / スペアパーツ

Nr
Order code
Specification