- F
- L
- E
- X
JT33 TDLAS gas analyzers with sample conditioning system enclosures (door open and closed)

- F
- L
- E
- X
JT33 TDLAS H2S gas analyzer with sample conditioning system enclosure, door closed

- F
- L
- E
- X
JT33 with a stainless-steel optical head, in a sample conditioning system enclosure with window

- F
- L
- E
- X
JT33 featuring solenoid valves with one-point validation, in a sample conditioning system enclosure box, door removed

- F
- L
- E
- X
JT33 featuring solenoid valves with two-point validation, in a sample conditioning system enclosure box, door removed

- F
- L
- E
- X
JT33 with a sample conditioning system enclosure (with window), mounted on a wall
Specs at a glance
-
Measuring principle
TDLAS
-
Analyte and measurement ranges
H2S (Hydrogen sulfide):
0 to 10 ppmv
0 to 500 ppmv
other ranges by request -
Hazardous area approvals
ATEX / IECEx /UKEx Zone 1
PESO / KTL / JPNEx Zone 1
INMETRO Zone 1
CNEx Zone 1
CSA Class I, Division 1
CSA Class I, Zone 1
Field of application
The JT33 gas analyzer offers continuous, real-time H2S measurement with low maintenance requirements. It enables plant operators and owners to meet gas quality standards, improve process control, and ensure asset integrity. Using proven differential technology, the JT33 tolerates stream changes while meeting measurement performance specifications. It also features auto-validation and advanced diagnostics supported by Heartbeat Technology for verification reporting.
Natural gas, biomethane, and carbon dioxide pipelines: prevent corrosion and minimize the risk of an uncontrollable event, ensuring asset integrity
Natural gas and LNG plants: optimize gas sweetening processes (amine, solid scavengers, and others)
NGL, refining, and petrochemical gas streams: assure gas quality and asset protection in fractionization, refining, and stream cracking processes
Carbon capture processes (CCUS): monitor processes and ensure that CO2 meets quality specifications
Benefits
Continuous, real-time TDLAS measurements with proven metrology to meet measurement specifications
Robust patented differential technology which tolerates contaminants and stream changes in tough applications
Auto-validation to confirm analyzer health in the field
Heartbeat Technology for automatically stored historical data, spectrum logging, diagnostics, and verification reporting for custody transfer applications
NIST-traceable factory calibration with superior accuracy and repeatability
Field-serviceable components and modules for minimal downtime and maximum repair flexibility
User-friendly interface with an intuitive menu and integrated web server software
Videos
-
The JT33 TDLAS gas analyzer product overview video
Xpert selection
-
Specialized products
- Designed for demanding applications
Technical excellence
Simplicity
Variable

-
Standard products
- Reliable, robust and low-maintenance
Technical excellence
Simplicity
-
Specialized products
- Designed for demanding applications
Technical excellence
Simplicity
Variable
FLEX selections | Technical excellence | Simplicity |
---|---|---|
Fundamental selection Meet your basic measurement needs |
Technical excellence
|
Simplicity
|
Lean selection Handle your core processes easily |
Technical excellence
|
Simplicity
|
Extended selection Optimize your processes with innovative technologies |
Technical excellence
|
Simplicity
|
Xpert selection Master your most challenging applications |
Technical excellence
|
Simplicity
Variable |
Success stories
-
Measuring principle
TDLAS
-
Product headline
The JT33 TDLAS H2S analyzer provides reliable, continuous, real-time H2S measurement with proven metrology and is distinguished by a robust differential measurement technology. It can monitor H2S in light or heavy streams, has a fast response to stream changes, and an ability to tolerate contamination with minimal maintenance for exceptional analyzer availability. It is supported by Heartbeat Technology for advanced diagnostics and utilizes superior measurement algorithms.
-
Channels
1
-
Analyte and measurement ranges
H2S (Hydrogen sulfide):
0 to 10 ppmv
0 to 500 ppmv
other ranges by request -
Measured variables
Concentration
Cell pressure
Cell temperature -
Ambient temperature range
JT33 TDLAS gas analyzer system:
–20 to 50 °C (–4 to 122 °F)
–10 to 60 °C (14 to 140 °F) -
Operating pressure range
Application dependent
800 to 1200 mbara (standard)
800 to 1700 mbara (optional) -
Analyzer wetted materials
316L stainless steel
FKM O-Rings
Glass
PCTFE/PTFE -
Power supply
For controller (spectrometer):
AC 100 to 240 V tolerance ±10 %, 50/60 Hz, 10W
DC 24 V tolerance ±20 %, 10W
UM = AC 250 V
For sample conditioning system electronics:
AC 100 to 240 V ±10 %, 50/60 Hz, 275W
UM = AC 250 V -
Communication
I/O1: Modbus RTU/RS485 or Modbus TCP/Ethernet
I/O2 and 3: Relay output or Universal I/O (UIO); UIO can be configured as analog input/output (4-20 mA) or digital/status output
Service web server interface: Ethernet RJ45 -
Housing materials
Electronics: Coated copper-free aluminum or Cast stainless steel
Sample system enclosure: 304 stainless steel or 316 stainless steel
Sample system panel: Anodized aluminum
Enclosure window (optional): Polycarbonate -
Hazardous area approvals
ATEX / IECEx /UKEx Zone 1
PESO / KTL / JPNEx Zone 1
INMETRO Zone 1
CNEx Zone 1
CSA Class I, Division 1
CSA Class I, Zone 1 -
Degree of protection
IP66, Type 4X
-
Product safety
CE
RCM
FCC
CRN
Choose from our suggested options
-
-- selected coding-- selected label99.999,00<%label%>
-
Looking for more options? Please use the individual configurator to continue
-
-- selected coding-- selected label99.999,00<%feature%>
-
<%label%>99.999,00
-
Looking for more options? Please use the individual configurator to continue
-
Process Connection
-
<%label%>99.999,00
-
<%label%>99.999,00
-
-
<%feature%><%coding%><%label%>
- CAD drawings CAD drawings
- CAD drawings
-
Prices per piece
-
<%range%><%price%>
-
- Optional add-ons
Individual configuration
We value your privacy
We use cookies to enhance your browsing experience, collect statistics to optimize site functionality, and deliver tailored advertisements or content.
By selecting "Accept all", you consent to our use of cookies.
For further details please review our cookie policy .